Determination of Dosimetric Characteristics of IrSeed 125I Brachytherapy Source

Message:
Abstract:
Introduction
Low dose rate brachytherapy sources have been widely used for interstitial implants in tumor sites, particularly in prostate. Dosimetric characteristics of a new IrSeed 125I brachytherapy source have been determined using the LiF thermoluminescent dosimeter (TLD) chips.
Materials And Methods
Dose rate constant, radial dose function, and anisotropy function around the IrSeed 125I source were measured in a plexiglass phantom using TLD-100 chips. A plexiglass slab phantom with dimensions of 30×30×7.3 cm3 was used to measure dose distribution around the source.
Results
Dose rate constant was measured to be equal to 0.965±0.006 cGyh-1U-1. Radial dose function, anisotropy function, and geometry function have been presented as tabulated data for the IrSeed source.
Conclusion
Basically, the dosimetric parameters presented here for this new IrSeed source have clinical and treatment planning applications.
Language:
English
Published:
Iranian Journal of Medical Physics, Volume:10 Issue: 2, Summer 2013
Pages:
109 to 117
magiran.com/p1229443  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!