New Potentiometric Sensor for Copper Determination in Waste Waters Based on a New Type of Modified GC Electrode

Message:
Abstract:
The potentiometrical analysis of Cu2+ using Nafion- modified electrode with Cu complex with DPABA (Methyl 3,5-bis{bis-[(pyridin-2-yl)methyl]amino}methyl)benzoate) incorporated, is described. Parameters and conditions such as the volume of 2% Nafion, pH, and the concentration of Cu complex with DPABA and pH of medium were optimized. Interference of Fe3+, Co2+ and Zn2+ with Cu2+ were examined. Calibration curve was linear in the range of 1×10-7-5×10-5 mol/L. The detection limit was 7×10-8 mol/L. A response time of the proposed electrode was 3 s.
Language:
English
Published:
Analytical & Bioanalytical Electrochemistry, Volume:4 Issue: 3, Jun 2012