Evaluation of the accordance of Radiation field with Light field in the Radiographic machines in Kerman X-Ray, Rafsanjan, and Sirjan cities, during 2010

Abstract:
Background And Objectives
The significance of the work entitle as the comparison of X-Ray field with light field is completely obvious for all of radiation workers. The accuracy and correctness of radiography image in duration of the performance of radiography has a direct connection with this issue. Unfortunately ihourcnuntry health system does not to the qualits consrup of radiography departments and patients are affected by the undesired radiations. Therefore, this study will focus on this issue in Kerman, Rafsanjan, and Sirjan.
Material And Methods
This study is an experimental-field which is performed on twenty-one active radiographic systems in Kerman, Rafsanjan and Sirjan cities during 2010. Due to the lack of access to the alignment test tool, a researcher's tool is made and testing is performed on cassette 24 × 30 centimeter in distance of 40 inches between the source of radiation source and cassette with the usage of light field with the size of 15 × 20 centimeter with the condition of exposure 50 KvP and 4 mAs, and the usage of small focal spot. It has recorded the exact information of evaluated system in the darkroom of exposed films and investigated the comparison of radiation X-Ray field with light field.
Results
This research evaluated 21 systems, 12 relevant systems in Kerman, 5 relevant systems in Rafsanjan, and 4 relevant systems in Sirjan. The average lack of comparison radiations field with light field in length and width cassette in Kerman in this manner were 1.09 % and 1.13% in Rafsanjan, 0.80% and 0.56% in Sirjan, 0.46 and 0.31%. The whole average with the discordance in length and width cassette in three mentioned city in this manner are 0.78% and 0.66%, respectively. Based on introduced protocol (2005) in Australia, and introduced protocol (2001) in Germany, the whole average of this result is acceptable but based on the introduced protocol (2005) in Australia, the average absence of comparison is not acceptable in Kerman and we should pay more attention to the case of Kerman.
Conclusion
With the attention to this point that Radiography is one of those diagnostic tool in medicine and it has the highest usage after the laboratory testing, it is essential to add some quality control program of radiography system to the most important duties of department of health physics at the medical science universities in Iran.
Language:
Persian
Published:
Community Health Journal, Volume:4 Issue: 3, 2010
Pages:
10 to 17
magiran.com/p1365768  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!