Comparative assessment of total and inhalable dust sampler performance in hexavalent chromium measurement

Message:
Abstract:
Background And Aim
Hexavalent chromium (Cr6+) is a known occupational carcinogen. Over the recent years, the increasing number of exposed workers and the emergence of serious health effects have led to considerable lowering of permissible exposure levels for Cr6+. Moreover, a number of methods have been developed for rapid and precise measurement of Cr6+concentrations. Given the importance of accurate measurements, this study was carried out to compare the performance of different samplers in relation to factors such as sampling height, sampling duration, and electroplating solution concentration.
Material And Methods
We used a hexavalent chromium generator fitted with a special sampling chamber and with proven ability to produce homogenous atmospheric concentrations. The performances of 4 samplers including IOM and CIS (as inhalable particle samplers) and Open and Closed-face cassettes (as total particle samplers) were examined through a simultaneous sampling technique. Sampling was done using the NIOSH 7600 method (samples were collected on PVC membrane filters with a pore size of 5.0 µm, at an air flow rate of 2.0 ± 0.1 l/min). Variables investigated with regard to their effects on sampler performance were sampling height (heights from bath solution: 35 and 50 cm), duration (30 and 180 min) and solution concentration (125 and 250 g/l).
Results
The measured concentrations ranged from 2 to 2725 µg.m-3. Analysis of the results in a completely randomized block design showed significant differences in performance between different samplers (P< 0.001). There were also differences between samplers at different heights and electroplating concentrations (P=0.0007). Multiple comparisons by Scheffe's method showed that the difference detected was due to the superior precision of the IOM (P<0.05).
Conclusion
Overally, there were statistically significant differences between the performances of the studied samplers; Cr6+ concentrations at the studied heights (4.09, 8.68 µg.m-3), sampling duration (5.90, 6.88 µg.m-3), and bath concentration (2.73, 11.88 µg.m-3) had no significant effects on sampler performance. The CIS sampler underwent corrosion during acidic mist collection and thus can not be recommended for Cr6+ mist collection.
Language:
Persian
Published:
Scientific Journal of School of Public Health and Institute of Public Health Research, Volume:5 Issue: 1, 2007
Pages:
47 to 55
magiran.com/p456318  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!