Compensation of Cross-Contamination in Simultaneous 201Tl/99mTc Myocardial Perfusion SPECT Imaging

Message:
Abstract:
Introduction
It is a common protocol to use 201Tl for the rest and 99mTc for the stress cardiac SPECT imaging. Theoretically, both types of imaging may be performed simultaneously using different energy windows for each radionuclide. However, a potential limitation is the cross-contamination of scattered photons from 99mTc and collimator X-rays into the 201Tl energy window. We used a middle energy window method to correct this cross-contamination.
Material And Methods
Using NCAT, a typical software torso phantom was generated. An extremely thin line source of 99mTc activity was placed inside the cardiac region of the phantom and no activity in the other parts. The SimSET Monte Carlo simulator was used to image the phantom in different energy windows. To find the relationship between projections in different energy windows, deconvolution theory was used. We investigated the ability of the suggested functions in three steps: Monte Carlo simulation, phantom experiment and clinical study. In the last step, SPECT images of eleven patients who had angiographic data were acquired in different energy windows. All of these images were compared by determining the contrast between a defect or left ventricle cavity and the myocardium.
Results
We found a new 2D kernel which had an exponential pattern with a much higher center. This function was used for modeling 99mTc down scatter distribution from the middle window image. X-ray distribution in the 201Tl window was also modeled as the 99mTc photopeak image convolved by a Gaussian function. Significant improvements in the contrasts of the simultaneous dual 201Tl images were found in each step before and after reconstruction. In comparison with other similar methods, better results were acquired using our suggested functions.
Conclusion
Our results showed contrast improvement in thallium images after correction, however, many other parameters should be evaluated for clinical approaches. There are many advantages in simultaneous dual isotope imaging. It halves imaging time and reduces patient waiting time and discomfort. Identical rest/stress registration of images also facilitates physicists’ motion or attenuation corrections and physicians’ image interpretation
Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:6 Issue: 24, 2010
Pages:
58 to 71
magiran.com/p722487  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!