A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Objective(s)
To investigate and compare quantitative accuracy of kidney absorbed dose measures made from both 2D and 3D imaging in patients receiving 177LuDOTATATE (Lutate) for treatment of neuroendocrine tumours (NETs).
Methods
Patients receiving Lutate therapy underwent both whole body planar imaging and SPECT/CT imaging over the kidneys at time points 0.5, 4, 24, and 96-120 hours after injection. Planar data were corrected for attenuation using transmission data, and were converted to units of absolute activity via two methods, using either a calibration standard in the field of view or relative to pre-voiding image total counts. Hand drawn regions of interest were used to generate time activity curves and kidney absorbed dose estimates in OLINDA-EXM. Fully quantitative SPECT data were generated using CT-derived corrections for both scatter and attenuation, before correction for dead time and application of a camera specific sensitivity factor to convert data to units of absolute activity. Volumes of interest were defined for kidney using the co-registered x-ray CT, before time activity curves and absorbed dose measures were generated in OLINDA-EXM, both with and without corrections made to the model for patient specific kidney volumes. Quantitative SPECT data were also used to derive dose maps through dose kernel convolution (DKC), which was treated as the gold standard.
Results
A total of 50 studies were analysed, corresponding to various cycles of treatment from 21 patients. Planar absorbed dose estimates were consistently higher than SPECT derived estimates by, on average, a factor of 3.
Conclusion
Quantitative SPECT is considered the gold standard approach for organ specific dosimetry however often relies on in house software. As such planar methods for estimating absorbed dose are much more widely available, and in particular, are often the only source of reference in previously published data. For the case of Lutate dosimetry, planar measures may lead to a three-fold increase in measures of kidney absorbed dose.
Language:
English
Published:
Asia Oceania Journal of Nuclear Medicine & Biology, Volume:6 Issue: 2, Spring 2018
Pages:
113 to 119
magiran.com/p1833595  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 990,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
دسترسی سراسری کاربران دانشگاه پیام نور!
اعضای هیئت علمی و دانشجویان دانشگاه پیام نور در سراسر کشور، در صورت ثبت نام با ایمیل دانشگاهی، تا پایان فروردین ماه 1403 به مقالات سایت دسترسی خواهند داشت!
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 50 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!