Validity and Reliability of COPD-6 Device for Detecting Chronic Obstructive Pulmonary Disease in High-Risk Individuals
Author(s):
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background
The current study was done to evaluate the validity and reliability of the Vitalograph COPD-6 portable device for detecting chronic obstructive pulmonary disease (COPD) in high-risk individuals in Iran.Materials and Methods
This research was a cross-sectional descriptive study. Forced expiratory volume in one second (FEV1), forced vital capacity (FVC), and FEV1/FVC using standard spirometer and FEV1, FEV6, and FEV1/FEV6 with COPD- 6 device were measured and recorded. Descriptive analysis was done.Results
19 of 122 patients (15.6%) were diagnosed with COPD. The COPD-6 had an acceptable performance for detecting COPD as assessed by the area under the receiver operating characteristic (ROC) curve (0.72 ; 95% CI: 0.42-0.86), with an average sensitivity of 84% and specificity of 98%, positive predictive value of 89%, and negative predictive value of 97%. The positive likelihood ratio resulted was 42 and the negative likelihood ratio was 0.16.Conclusion
COPD-6 is a validate and reliable device for detecting COPD in non-specialized health care settings and the best cut-off point for FEV1/FEV6 ratio is 0.72.Keywords:
Language:
English
Published:
Tanaffos Respiration Journal, Volume:19 Issue: 3, Summer 2020
Pages:
201 to 207
https://magiran.com/p2237471
مقالات دیگری از این نویسنده (گان)
-
The Level of Community Readiness in Isfahan University of Medical Sciences Regarding the Reopening for in Person Training During the COVID-19 Pandemic
, Reza Khadivi *, Shima Safazadeh, Jalal Karimi
Journal Of Isfahan Medical School, -
Risk Factors for the Mortality in Hospitalized Patients with COVID-19: A Brief Report
Ramin Sami, MohammadReza Hajian, Babak Amra, Forogh Soltaninejad, Marjan Mansourian, Sam Mirfendereski, , Nilufar Khademi, Soheila Jalali, Nafiseh Shokri Mashhadi *
Iranian Journal of Medical Sciences, Nov 2021