Undesired Outcomes of the Catania Stent Compared to the Xience Stent in Patients Undergoing Angioplasty: A Double-Blind Randomized Controlled Trial
The present study tries to compare the unintended outcomes of the Catania stent versus Xience stent in patients undergoing angioplasty.
In a three month, follow-up, double-blinded, randomized controlled trial, 83 patients undergoing angioplasty, who met the inclusion criteria were entered into the study. After randomization 43 patients were treated with the Xience stent and 40 patients with the Catania stent. Stent-related outcomes such as Cardiac and Non-Cardiovascular Death, Myocardial Infarction (MI), Target Lesion Revascularization (TLR), Stent Thrombosis (ST), Coronary Artery Bypass Grafting (CABG), Peripheral vasculopathy, and Cerebral Vascular Accident (CVA) were compared between the groups.
There was no statistically signifi cant difference in the incidence rate of complications and clinical outcomes between the two treatment groups (P > 0.05). The incidence of MI, TLR, CABG operation, peripheral vasculopathy, or CVA was not observed in any patient and there was no statistically difference in mortality (4.7% vs. 2.5%; P = 0.527) and stent thrombosis (2.3% vs. 2.5%; P = 0.735).
All in all, the present study could not fi nd the signifi cant differences between the Catania stent and Xience stent in terms of clinical outcomes during the follow-up period.