Background and Aim

 Dental implants are one of the most accepted methods for tooth replacement, function and esthetic improvement. Sinabionic dental implant (Nikkasht Asia, Iran) was manufactured by grade 5 titanium alloy with Sandblasted, large grit, Acid-etched (SLA) and hydrophilic surface with a porosity of 1.2 to 1.8 µm. The aim of this study was clinical and radiological evaluation of Sinabionic implant system at 3, 6 and 12 month intervals after implant placement surgery.

Material and Methods

 After obtaining written consent from the patients, 53 Sinabionic implants (Niktash Asia, Iran) were evaluated in this study.  Pocket depth, bleeding on probing, pus, mobility and pain were evaluated in clinical examination and bone resorption in mesial and distal were assessed in the radiographic examination by one sample t test. 

Results

 5 failures were observed of 53 implants and 32 implants were successful according to the accepted criteria, which includes 60.4% and the Survival rate of these implants was reported to be 90.6%.

Conclusion

  Clinical and radiographic evaluations of Sinabionic implants indicated success and survival rate less than the accepted standard range. It is strongly recommended that the structure (micro and macro geometry) and surface quality of the implant be modified.