Absolute Activity Measurement of 125I Seed Brachytherapy by Two-Dimensional Photon-Photon Coincidence Spectroscopy

Message:
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:

Photon-photon coincidence counting is a primary standardization technique used for the absolute activity measurement of short half-life radionuclides. In this study, a coincidence spectrometer based on digital signal processing was developed for the absolute standardization of 125I seed Brachytherapy. The spectrometer includes a waveform digitizer that samples at the preamplifier signals directly from two similar 2ʺ × 2ʺ NaI (Tl) detectors. The sampled signals were shaped into trapezoidal signals for pulse height analysis. These detected signals were then recorded in a list file and the software was used to recognize coincidence events. Detailed discussions were included on the formulas for determining absolute activity measurement. The activity of a 125I seed was measured and showed good agreement with the reference values. This presented technique is a quick method for measuring the activity of 125I at nuclear hospitals. The standardized source can also serve as a tracer for measuring the activity of the 109Cd source

Language:
English
Published:
Journal of Nuclear Research and Applications, Volume:4 Issue: 2, Spring 2024
Pages:
29 to 34
https://magiran.com/p2747336