The aim of this study was to report the results of using PDA occluders (coil occluder and Amplatzer device), which has continued since December 1999 at the pediatric department of Rajaee Heart Center. Transcatheter Occlusion of PDA M. Meraji MD, et al.
Between December 1999 and September 2002, 193 cases of PDA were admitted at the pediatric ward. Seventy cases underwent transcatheter occlusion of PDA by the detachable coil occluder, and in 53 cases PDA was occluded by the Amplatzer device.
Successful coil placement was accomplished in 68 cases. Two cases of intravascular hemolysis were observed due to residual shunt .In one of them, acute tubular necrosis ensued, and peritoneal dialysis was performed. However, after surgical PDA closure, the renal performance was recovered eventually. Also, two cases of coil detachment to the LPA were observed with no long-term sequala on pulmonary function. Fifty-two of the 53 Amplatzer occlusions were successful. Only one failure was observed due to the small size of the Amplatzer in comparison to the duct diameter.
Retrograde and antegrade transcatheter closure of the PDA by the detachable coil occluder and Amplatzer device has been performed successfully in our department. The problem that we face has been PDA sizing and also the retrieval of detached coils
Between December 1999 and September 2002, 193 cases of PDA were admitted at the pediatric ward. Seventy cases underwent transcatheter occlusion of PDA by the detachable coil occluder, and in 53 cases PDA was occluded by the Amplatzer device.
Successful coil placement was accomplished in 68 cases. Two cases of intravascular hemolysis were observed due to residual shunt .In one of them, acute tubular necrosis ensued, and peritoneal dialysis was performed. However, after surgical PDA closure, the renal performance was recovered eventually. Also, two cases of coil detachment to the LPA were observed with no long-term sequala on pulmonary function. Fifty-two of the 53 Amplatzer occlusions were successful. Only one failure was observed due to the small size of the Amplatzer in comparison to the duct diameter.
Retrograde and antegrade transcatheter closure of the PDA by the detachable coil occluder and Amplatzer device has been performed successfully in our department. The problem that we face has been PDA sizing and also the retrieval of detached coils
Keywords:
Language:
English
Published:
Iranian Heart Journal, Volume:6 Issue: 1, Spring-Summer 2005
Page:
43
magiran.com/p610161
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یکساله به مبلغ 1,390,000ريال میتوانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
- حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران میشود.
- پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانههای چاپی و دیجیتال را به کاربر نمیدهد.
In order to view content subscription is required
Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!