Estimation and assessment of the effective doses for radiosensitive organs in women undergoing chest CT scans with or without automatic exposure control system

Message:
Abstract:
Background
There are several techniques for reducing the delivered dose from CT (Computed tomography) scanning such as the automatic exposure control (AEC). This technique modulates the tube current regarding the patient size and weight. The aim of this study was to estimate the effect of the AEC on the radiosensitive organs effective doses in women undergoing chest CT scanning.
Methods
This study was a cross-sectional, analytical and quantitative study that was performed during 3 months in the imaging section of the Firoozgar educational and therapeutic hospital (belonging to Shahid Beheshti University of Medical Sciences) in the spring of 2017. CT scan exposure parameters were gathered and registered for 54 women undergoing chest CT scan. 25 of these scans were performed using AEC system and 29 of them were performed without using AEC. CT dose indexes in the center and peripheral regions of the standard phantom were calculated using the exposure parameters. Weighted CT dose index was also calculated and effective organ doses were obtained using CT-Expo software, version 2 (Medizinische Hochschule, Hannover, Germany) for two mentioned groups. In addition, noise was measured for these two groups as an image quality parameter.
Results
Calculated weighted CT dose indexes were 9.94 mGy and 12.46 mGy using AEC system and without using AEC, respectively. The calculated effective doses were equal to 5.4 mSv and 6.3 mSv using AEC and without using AEC, respectively. Maximum organ effective doses were 15, 14, 14 and 14 mSv for breast, esophagus, lung and thymus respectively in the non-using AEC system imaging technique.
Conclusion
Our measurements indicated a decrease about 15% in weighted CT dose index (from 12.46 to 9.94 mGy) using AEC system. Beside of this fact, the noise increased about 11.3% (from 4.2 to 4.74) using AEC system. So, it can be said that using of AEC was an effective way for dose reduction in women undergoing chest CT scanning, and the additional noise was in the acceptable range.
Language:
Persian
Published:
Tehran University Medical Journal, Volume:75 Issue: 7, 2017
Pages:
496 to 503
magiran.com/p1760632  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با عضویت و پرداخت آنلاین حق اشتراک یک‌ساله به مبلغ 1,390,000ريال می‌توانید 70 عنوان مطلب دانلود کنید!
اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!