T1 Thermometry for Deep Brain Stimulation Applications: A Comparison between Rapid Gradient Echo Sequences
Author(s):
Article Type:
Research/Original Article (دارای رتبه معتبر)
Abstract:
Background
T1 thermometry is considered a straight method for the safety monitoring of patients with deep brain stimulation (DBS) electrodes against radiofrequency-induced heating during Magnetic Resonance Imaging (MRI), requiring different sequences and methods.Objective
This study aimed to compare two T1 thermometry methods and two low specific absorption rate (SAR) imaging sequences in terms of the output image quality.Material and Methods
In this experimental study, a gel phantom was prepared, resembling the brain tissue properties with a copper wire inside. Two types of rapid gradient echo sequences, namely radiofrequency-spoiled and balanced steady-state free precession (bSSFP) sequences, were used. T1 thermometry was performed by either T1-weighted images with a high SAR sequence to increase heating around the wire or T1 mapping methods.Results
The balanced steady-state free precession (bSSFP) sequence provided higher image quality in terms of spatial resolution (1×1×1.5 mm3 compared with 1×1×3 mm3) at a shorter acquisition time. The susceptibility artifact was also less pronounced for the bSSFP sequence compared with the radiofrequency-spoiled sequence. A temperature increase, of up to 8 ℃, was estimated using a high SAR sequence. The estimated change in temperature was reduced when using the T1 mapping method. Conclusion
Heating induced during MRI of implanted electrodes could be estimated using high-resolution T1 maps obtained from inversion recovery bSSFP sequence. Such a method gives a direct estimation of heating during the imaging sequence, which is highly desirable for safe MRI of DBS patients.Keywords:
Language:
English
Published:
Journal of Biomedical Physics & Engineering, Volume:14 Issue: 6, Nov-Dec 2024
Pages:
569 to 578
https://magiran.com/p2794495