CT-number to Electron Density Calibration Using Stoichiometry for Use in Radiotherapy

Message:
Abstract:
Introduction
One of the causes of error and perturbation in isodose curves in conventional radiotherapy treatment planning is the existence of tissues having either very high density (bone, prosthesis) or very low density (lung, air cavities). Nowadays, the use of CT images to solve this problem is growing, and so is the need for calibration curves that convert the CT-number of the tissue to electron density for dose calculation. The conventional method to obtain this curve is purely measurement-based, in which a phantom containing various materials of known electron densities is imaged. Alternatively, a more fundamental method of stoichiometry has been used in this work.
Material And Methods
For the stoichiometric method, initially a cylindrical polyethylene phantom was built. The phantom consists of inserts of high-purity aluminum, PVC, polyethylene, water and cork to model hard bone, skeleton, fat, muscle and lung tissues, respectively. CT imaging was then performed at 120 kVp using a spiral CT scanner (GE model NXI). A system of simultaneous equations was solved to get the appropriate CT-number to electron-density conversion for each tissue type using the CT-numbers from the phantom images, the physical and radiological data of the materials. A conversion curve showing the variation of CT-number with relative electron density was also plotted. The result of the stoichiometric conversion was then compared to that from other methods.
Results
The system of simultaneous equations yielded the factors, ,. The experimental and the computed CT-numbers were compared. The highest uncertainty was estimated to be approximately 5.6% for a relatively high-density material such as aluminum and 4% for polyethylene. The curves representing the electron density based on CT-number start diverging at CT-number equal to zero and above.Discussion and
Conclusion
Comparing the results obtained from the experimental and computational methods suggest an acceptable level of accuracy for the computational (stoichiometric) conversion. The uncertainty in the electron density obtained is greater for materials of higher electron density.
Language:
Persian
Published:
Iranian Journal of Medical Physics, Volume:3 Issue: 13, 2007
Pages:
59 to 66
https://www.magiran.com/p511371  
دانلود و مطالعه متن این مقاله با یکی از روشهای زیر امکان پذیر است:
اشتراک شخصی
با ثبت ایمیلتان و پرداخت حق اشتراک سالانه به مبلغ 1,390,000ريال، بلافاصله متن این مقاله را دریافت کنید.اعتبار دانلود 70 مقاله نیز در حساب کاربری شما لحاظ خواهد شد.

پرداخت حق اشتراک به معنای پذیرش "شرایط خدمات" پایگاه مگیران از سوی شماست.

اگر مقاله ای از شما در مگیران نمایه شده، برای استفاده از اعتبار اهدایی سامانه نویسندگان با ایمیل منتشرشده ثبت نام کنید. ثبت نام

اشتراک سازمانی
به کتابخانه دانشگاه یا محل کار خود پیشنهاد کنید تا اشتراک سازمانی این پایگاه را برای دسترسی نامحدود همه کاربران به متن مطالب تهیه نمایند!
توجه!
  • حق عضویت دریافتی صرف حمایت از نشریات عضو و نگهداری، تکمیل و توسعه مگیران می‌شود.
  • پرداخت حق اشتراک و دانلود مقالات اجازه بازنشر آن در سایر رسانه‌های چاپی و دیجیتال را به کاربر نمی‌دهد.
In order to view content subscription is required

Personal subscription
Subscribe magiran.com for 70 € euros via PayPal and download 70 articles during a year.
Organization subscription
Please contact us to subscribe your university or library for unlimited access!